Optical Fibre Cable Testing Equipment (OFC Testers) - Optical Cable Flammability Tester - Flame Propagation Tester for a Single Insulated Cable
Optical Cable Flammability Tester - Flame Propagation Tester for a Single Insulated Cable
IEC 60332-1-1, IEC 60332-3-10/21/22/24/25
Optical Cable Flammability Tester - Flame Propagation Tester for a Single Insulated Cable TTOFTFB-2000 is designed to evaluate the flame resistance of optical cables.
Test sample
The test sample must consist of multiple cable segments from the same production batch, each at least 3.5 meters long.
The total number of segments should ensure a nominal combined volume of 7 liters per meter of non-metallic material.
Before testing, condition the cable segments at a temperature of 20 ± 10°C for a minimum of 16 hours. The segments must be dry.
Determination of the number of test pieces
To calculate the required number of test pieces, determine the volume per meter of non-metallic material in a single test piece.
Carefully cut a cable segment at least 0.3 meters long, ensuring the cuts are perpendicular to the cable axis for accurate length measurements.
Flame application time
Apply the test flame for 40 minutes, then extinguish it. Maintain the airflow in the test chamber until the cable stops burning or glowing, or for a maximum of 1 hour. After this period, extinguish any remaining burning or glowing cable.
Evaluation of test results
After all cable burning or glowing has stopped or been extinguished, wipe the test sample clean.
Disregard soot if the original surface remains undamaged after cleaning. Ignore any softening or deformation of the non-metallic material.
Measure flame spread as the length of the damaged area in meters, accurate to two decimal places, from the bottom edge of the burner to the point where charring begins. To identify the onset of char, press a sharp object (such as a knife blade) against the cable surface. The transition from a resilient to a brittle, crumbling surface indicates the onset of char.
Performance requirements
Performance requirements for a specific type or class of wire or cable should ideally be specified in the respective cable standard. If no requirements are provided, refer to the recommended performance criteria outlined in Annex B.
Model | TTOFTFB-2000 |
---|---|
Chamber Size | W1000 x D2000 x H4000mm |
Inlet Air Flow Rate | 5000 +/- 500L/min |
Fuel Type | Propane |
Combustion-supporting gas | Compressed Air > 1MPa |
Ignition | Automatic |
Exhaust Fumes Filtering Tank | Filter exhaust fumes throughout the test |
Testing chamber | Schematic for the chamber to be manufactured locally |
Voltage | AC380V 50Hz 1kW |