Face Mask Clogging Tester
Face Mask Clogging Tester ToronFM-CT – The sample is exposed to a known concentration of dolomite dust in the air. After exposure, the breathing resistance and filter penetration of the particle-filtering half mask are measured.
Standards
- EN 149 section 8.10
Parameter | Specification |
---|---|
Model | ToronFM-CT |
Dust | DRB 4/15 dolomite |
Dust Distributor | Particle size range: 0.1 µm ~ 10 µm Mass Flow Range: 40mg/h --- 400mg/h |
Breathing Machine | Displacement: 2.0L/stroke Frequency: 15 times/min |
Exhaled Air Temperature | (37 ± 2)°C |
Exhaled Air Humidity | 95% R.H minimum |
Continuous Flow Through the Dust Chamber | 60 m³/h, linear velocity: 4 cm/s |
Concentration of the Dust | (400 ± 100) mg/m³ |
Test Chamber | Inner Dimension: 650 × 650 mm × 700 mm Airflow: 60 m³/h, linear velocity: 4 cm/s Air Temperature: (23 ± 2)°C Air Humidity: (45 ± 15)% |
Respiratory Resistance Measurement Range | 0 ~ 2000 Pa, accuracy: 0.1 Pa |
Voltage | 220V, 50Hz, 1KW (110V is also available) |
External Dimension | 3800mm × 1100mm × 1650mm |
Weight | 320kg |
Accessories
Item | Quantity |
---|---|
Automatic dolomite dust generator | 1pc |
Automatic powder feeding center system (Continuous feeding device) | 1set |
TSI laser dust particle counter | 1pc |
Sinusoidal wave simulates ventilator | 1set |
Saturated steam generator | 1set |
Please Contact us for pricing, check availability, or ask for additional information about our Face Mask Clogging Tester